Comparison of spectral domain optical coherence tomography and ultrasonic pachymetry for assessment of central corneal thickness

Küçük Resim Yok

Tarih

2014

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Turkish Ophthalmology Society

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

Objectives: The aim of this study was to determine if there is a difference in central corneal thickness (CCT) measurements obtained by Cirrus spectral domain optical coherence tomography (SD-OCT) and ultrasonic pachymetry in healthy individuals. Materials and Methods: The study included 50 healthy consecutively selected individuals without ocular or systemic disease. CCT was first measured using OCT, and then using ultrasonic pachymetry. Results: Mean age of the participants was 31.44 years. Mean CCT measured using SD-OCT was 531.78 µm versus 535.15 µm by ultrasonic pachymetry. Mean CCT measurement obtained by Cirrus SD-OCT showed statistically significant difference by approximately 3.37 µm than the one obtained by ultrasonic pachymetry (t-test, p<0.05); however, Bland-Altman analysis proved that there was high concordance between the measurements. Conclusion: CCT measurements obtained by Cirrus SD-OCT were very similar to those obtained by ultrasonic pachymetry, and as such we think that Cirrus SD-OCT can be used in our present ophthalmology practice to measure OCT.

Açıklama

Anahtar Kelimeler

Central corneal thickness, Spectral domain optical coherence tomography, Ultrasonic pachymetry

Kaynak

Turk Oftalmoloiji Dergisi

WoS Q Değeri

Scopus Q Değeri

Cilt

44

Sayı

4

Künye