Trace enrichment and atomic absorption spectrometric determination of lead, copper, cadmium and nickel in drinking water samples by use of an activated carbon column

dc.authorid0000-0002-1017-0244
dc.contributor.authorSoylak, M
dc.contributor.authorNarin, I
dc.contributor.authorDogan, M
dc.date.accessioned2019-08-01T13:38:39Z
dc.date.available2019-08-01T13:38:39Z
dc.date.issued1997
dc.departmentNiğde ÖHÜ
dc.description.abstractA simple method for atomic absorption spectrometric determination of lead copper, cadmium and nickel in drinking water samples after preconcentration by sorbing 1-(2-pyridylazo) 2-naphthol (PAN) complex of these metals on an activated carbon column has been established. The metal/PAN complexes were quantitatively retained on the activated carbon in the pH range 6-8. Metals retained on the activated carbon column were completely eluted with 2M HCl in acetone. This method was applied to the determination of lead, copper, cadmium and nickel in drinking water samples and good results were obtained (Recoveries >95%, relative standard deviations <7%, relative error <3%).
dc.identifier.doi10.1080/00032719708001823
dc.identifier.endpage2810
dc.identifier.issn0003-2719
dc.identifier.issue15
dc.identifier.scopus2-s2.0-0030706161
dc.identifier.scopusqualityQ3
dc.identifier.startpage2801
dc.identifier.urihttps://dx.doi.org/10.1080/00032719708001823
dc.identifier.urihttps://hdl.handle.net/11480/5835
dc.identifier.volume30
dc.identifier.wosWOS:A1997YK02900013
dc.identifier.wosqualityQ3
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.institutionauthor[0-Belirlenecek]
dc.language.isoen
dc.publisherMARCEL DEKKER INC
dc.relation.ispartofANALYTICAL LETTERS
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectactivated carbon column
dc.subjectcopper
dc.subjectlead
dc.subjectcadmium
dc.subjectnickel
dc.subject1-(2-pyridylazo) 2-naphthol
dc.subjectflame AAS
dc.titleTrace enrichment and atomic absorption spectrometric determination of lead, copper, cadmium and nickel in drinking water samples by use of an activated carbon column