Auditory Threshold Detection by Classifying Estimated Short Latency Evoked Potentials

Küçük Resim Yok

Tarih

2009

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

IEEE

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Short latency evoked response (SLER) has become a routine clinical tool in neurological and audiological assessment. But, in order to extract SLER from backgroung EEG signal, many repeated single trial measurements are necessary In some cases, these reprtitions are up to 2000. Therefore, measuring period is very time consuming and uncomfortable for subjects. This condition is also limited the SLER usage in clinical applications. In this study, 302 SLER responses extracted by averaging 1024 single trials are used for creating two different data sets. The first set is created from ensemble averaging of 1024 trials for each SLER signals. The second set is obtained from the same single trial measurements by estimating 64 trials of each SLER signal. The support vector machine which is a powerful binary classifier is performed for each data sets for three different feature extraction techniques. In result, the results obtained from estimated data (second data set) classification procedure is better than the results of classical ensemble averaged data set (first data set) with a high accuracy and less time consuming. This results contribute to the SLER usage in clinics more practical than classical ones.

Açıklama

14th National Biomedical Engineering Meeting -- MAY 20-22, 2009 -- Izmir, TURKEY

Anahtar Kelimeler

Kaynak

BIYOMUT: 2009 14TH NATIONAL BIOMEDICAL ENGINEERING MEETING

WoS Q Değeri

N/A

Scopus Q Değeri

N/A

Cilt

Sayı

Künye