XAD-4/PAN solid phase extraction system for atomic absorption spectrometric determinations of some trace metals in environmental samples

Küçük Resim Yok

Tarih

2004

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

TAYLOR & FRANCIS INC

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

A simple, fast, and sensitive column solid phase extraction study has been proposed for flame atomic absorption spectrometric determination of lead, cadmium, chromium, nickel, copper, and manganese ions on XAD-4 as 1-(2-pyridylazo)2-naphthol (PAN) complexes. The analytes were sorbed as PAN complexes on Amberlite XAD-4 column at the pH range of 8.5-10, and then eluted with 1 M HNO3 in acetone. The effect of pH, reagent amount, sample volume, and matrix of the sample on the sorption of metal ions has also been investigated. The concentration of analytes after separation and preconcentration was in agreement with the added amount of ions. The present method was found to be applicable to the preconcentration of analyte ions in river water from Yesilirmak, lake water from Lake Van, bottled mineral waters, tap water, stream sediment, mushroom, tobacco, and black tea samples with good results (recoveries greater than 95%, relative standard deviations lower than 10%).

Açıklama

Anahtar Kelimeler

enrichment, amberlite XAD-4, column, trace metal, FAAS

Kaynak

ANALYTICAL LETTERS

WoS Q Değeri

Q3

Scopus Q Değeri

Q3

Cilt

37

Sayı

3

Künye